Summary of EDI Damage Causes

2023-05-06 09:55:11   Visit:863

The service life of EDI membrane stack is limited. Whether it is imported or domestic, we all know that under normal use, the service life of EDI is approximately 3-5 years. In the process of EDI maintenance, cases of failure within half a year or one year are often received, which are not limited to imported and domestic brands. So what are the causes of EDI module damage?

Here are some reasons we often encounter during maintenance:

永旺彩票1、 EDI does not produce water or does not produce water.

① The residual chlorine in the influent water exceeds the standard, resulting in the breakage of EDI resin, blocking the channel, and reducing the water yield;

永旺彩票② The module has not been cleaned and maintained for a long time, resulting in the EDI internal structure;

永旺彩票③ Water shortage operation directly leads to module dry burning deformation;

2、 The EDI operating voltage is very high and the current cannot be adjusted.

永旺彩票① Scaling inside EDI leads to increased voltage and decreased water quality and volume;

② Water hammer phenomenon occurs during equipment operation, resulting in uneven resin distribution within EDI and voltage rise;

3、 EDI produced water resistivity is not up to standard.

永旺彩票① The inlet conductivity is not up to standard, and the inlet conductivity is too high;

② The PH value of influent water is not within the appropriate range;

永旺彩票③ Whether the EDI operating current is turned on too low or too high;

In response to the above issues, we should pay attention to the following points during normal use:

永旺彩票1、 Intake water quality assurance

永旺彩票1. Check the incoming water quality, install an incoming water conductivity meter on the equipment, and regularly monitor the conductivity of the incoming water quality (check whether the instrument probe is abnormal);

2. It is also necessary to regularly monitor the residual chlorine in the water, and residual chlorine testing agents can be used for testing. It is recommended to increase oxidation reduction treatment or replace activated carbon if the residual chlorine exceeds the standard (excessive residual chlorine or activated carbon adsorption fails)

永旺彩票3. It is recommended to maintain the PH value of the incoming water within the range of 6.5-9, and the water quality should preferably be slightly alkaline;

2、 Equipment operation guarantee

永旺彩票1. Regularly clean and maintain EDI modules to ensure stable operation of the equipment;

2. It is recommended to separate the EDI module host. Although the integration is beautiful, the system stability is relatively not so good;

永旺彩票3. Match EDI modules of different tonnage to the appropriate operating current